R36m clinical tympanograph RESONANCE Italy
The clinical tympanograph performs: Tympanometry, Acoustic Reflex, Reflex Decay, Quick test: check, screening and Decay, Acoustic Reflex Latency test (ARLT), Multi-frequency Tympanometry, ETF (Intact perforated and patulous), Special tests (Growth-DLI and Non-acoustic).
Recordings
· It has a built-in thermal printer
· A pdf report is created directly from the device and saved to USB, with the ability to add patient details and comments via a wireless keyboard (optional)
· It has pediatric audiometry
· Power supply 110-240 V AC 50/60 Hz 40VA
· It has dimensions of 370x290x180mm and weighs 3.5 kg.
· ISO 13485, ISO 9001, MDD 93/42/EEC, Class IIa, IEC 60601-1, IEC 60601-1-2, IEC 60645-5/ ANSI S.3.39 with CE and FDA approval.
Tympanometry features
· Intensity: 226Hz: 85dB SPL +-2dB 678, 800 and 1000Hz: 75dB SPL +-2dB Frequency accuracy +-0.5%
· Range at 226Hz: 0.05 to 7ml
· Sensitivity scale: Automatically adjusts to the appropriate range, available scales at 226Hz: 1.5, 2, 5, 7 ml
· Range at 678, 800, 1000Hz: 0 to +25mmho
· HF sensitivity scale: Automatically adjusts to the appropriate range, available scales at 678, 800 and 1000Hz: 5,10, 15, 20, 25 mmho
· Pressure adjustment automatically and manually
· Range from +400 to -600daPa and from +200 to -400daPa, both adjustable in steps of 50daPa
· Pressure accuracy +-10daPa or +-10%
· Ρυθμός σάρωσης: 50,100,200, 300daPa/sec και αυτόματα
· Indicator: The measured value is displayed
· Protection limits: -800 to +600daPa
· Eustachian tube control function
· ETF test for intact and perforated eardrums
· Also available is a test for a pulsatile tympanic membrane
· Reflex testing methods: Threshold, automatic and manual
· Stimulus duration: 0.5, 1, 2 seconds
· Ability to select protocols for all examination methods
· Ipsilateral and contralateral stimuli for all reflex tests
· Automatic peak search functions for all examination methods
· Manual Reflex: Manual control of all stimuli
· Reflex Decay: Με κατώφλι, αυτόματα και χειροκίνητα
· Stimulus duration 10 or 20 seconds
· ARLT with threshold, automatic and manual
· Stimulus duration: 1 second
Ipsilateral reflexes
- Tone level range (dB HL) from 50 to 110
- Noise level range (dB SPL) from 50 to 100
- Frequency: 500, 1000, 2000, 3000, 4000 Hz
- Noise: BBN, HP, LP
- Frequency accuracy: +-1%
- Harmonic distortion (THD): less than 3%
Unilateral reflexes
- Tone level range (dB HL) from 50 to 120
- Noise level range (dB SPL) from 50 to 115
- Frequency: 250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz
- Noise: BBN, HP, LP
- Frequency accuracy: +-1%
- Calibration accuracy: +-3 dB
- Steps: 1,2,5,10 dB
- On/off ratio 70dB minimum
Includes:
- 226Hz or HF nozzle
- Headband
- Nozzle handle
- Nozzle cleaning kit
- DD45 for Unilateral examination
- Entrance for contralateral
- Kit with nipples of various sizes from 6 mm to 15 mm.
- Calibration well with swab holder
- Built-in thermal printer
- 1 roll of paper with a width of 112 mm.
- Power cord (110-220 V)
- Protective cover
- User Manual
- Resonance MDS software including NoAH (demo version)
- 1 extra fuse